User Tools

Site Tools


chara:operating_procedures

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision Both sides next revision
chara:operating_procedures [2018/07/05 08:26]
jones [Lab Protocol]
chara:operating_procedures [2018/07/05 08:31]
jones [Preparations in the Lab Prior to Observing]
Line 84: Line 84:
 \\  \\ 
 \\  \\ 
 +**Chapter 2:** Version 2. August 27, 2012 by Judit, Version 3 January 2015 by Norm, Version ? October 2017 by Norm\\
 ====Preparations in the Lab Prior to Observing==== ====Preparations in the Lab Prior to Observing====
-\\  
-Version 2. August 27, 2012 by Judit, Version 3 January 2015 by Norm, Version ? October 2017 by Norm\\ 
 \\  \\ 
 These are routine tasks to be performed in and around the lab every evening to prepare the array for regular observations.\\  These are routine tasks to be performed in and around the lab every evening to prepare the array for regular observations.\\ 
 \\  \\ 
 ===2.1 Starting the vacuum pump for the light pipes=== ===2.1 Starting the vacuum pump for the light pipes===
-\\  
 Go into vacuum pump shack\\  Go into vacuum pump shack\\ 
  
Line 106: Line 104:
 \\  \\ 
 ===2.2 Filling IR cameras with LN2=== ===2.2 Filling IR cameras with LN2===
-\\  
 \\  \\ 
 Fill both chambers on the NIRO camera. Use the gloves and glasses provided on the shelf with the funnels and the thermos. It will normally take 1 liter of LN2 in the outer chamber and 1/2 liter in the inner chamber. Fill it until the LN2 bubbles out of the top. This camera is normally kept cool all the time, exceptions will be noted.\\  Fill both chambers on the NIRO camera. Use the gloves and glasses provided on the shelf with the funnels and the thermos. It will normally take 1 liter of LN2 in the outer chamber and 1/2 liter in the inner chamber. Fill it until the LN2 bubbles out of the top. This camera is normally kept cool all the time, exceptions will be noted.\\ 
Line 115: Line 112:
 \\  \\ 
 ===2.3 Alignment of the light path to the telescopes=== ===2.3 Alignment of the light path to the telescopes===
-\\  
 \\  \\ 
 Go into the lab. The green alignment laser on the light source table is used. To open necessary guis, use laptops or the two tablets inside lab. Tablets are the newest hand held units and are located on the west and south walls plugged into chargers.\\  Go into the lab. The green alignment laser on the light source table is used. To open necessary guis, use laptops or the two tablets inside lab. Tablets are the newest hand held units and are located on the west and south walls plugged into chargers.\\ 
Line 199: Line 195:
 \\  \\ 
 ===2.5 Starting up OPLE and Metrology=== ===2.5 Starting up OPLE and Metrology===
-\\  
 \\  \\ 
 Go to OPLE racks\\  Go to OPLE racks\\ 
Line 227: Line 222:
 \\  \\ 
 ===2.6 Final steps in the lab=== ===2.6 Final steps in the lab===
-\\  
 \\  \\ 
  
chara/operating_procedures.txt ยท Last modified: 2023/09/01 02:50 by gail_stargazer