1.2 The Metrology Laser | 1.2 The Metrology Laser | ||
- | A high-powered infrared laser is used for the delay line metrology. This is an eye hazard, so before entering the lab check to make sure the red light above the door is not on. If the laser is activated, it is a must to put on IR-protective goggles. If you are not sure, wear the goggles. Please note that the METLAS gui or OPLE controller | + | A high-powered infrared laser is used for the delay line metrology. This is an eye hazard, so before entering the lab check to make sure the red light above the door is not on. If the laser is activated, it is a must to put on IR-protective goggles. If you are not sure, wear the goggles. Please note that the METLAS gui or OPLESystem Controller |
1.3 General Lab Rules | 1.3 General Lab Rules | ||
Line 94: | Line 96: | ||
Go into vacuum pump shack. Flip on power switch for pump and wait for the blower to go on in about 6 seconds. Open the valve slowly after about 10-12 seconds (parallel to pipe means open). Note if fan cords are plugged into wall outlet to keep pumps cool in the summer. Note oil temperature when you start up. Oil temps are usually kept above 15º C by a block heater on the north side of the pump and a space heater on the south side. Inform Larry or Craig if the oil temp is low and the heaters did not run to bring the temp up before starting.\\ | Go into vacuum pump shack. Flip on power switch for pump and wait for the blower to go on in about 6 seconds. Open the valve slowly after about 10-12 seconds (parallel to pipe means open). Note if fan cords are plugged into wall outlet to keep pumps cool in the summer. Note oil temperature when you start up. Oil temps are usually kept above 15º C by a block heater on the north side of the pump and a space heater on the south side. Inform Larry or Craig if the oil temp is low and the heaters did not run to bring the temp up before starting.\\ | ||
- | Go into lab building, check pressures on Vacmon display. If pressure is higher than 100 torr in any of the lines, pump those lines only one at a time. Open each valve slowly, you will hear the air rush out as the pump pulls the air. Slowly continue to fully open. Open each valve as the pressure is pumped down to the pressure in the next lowest line. Ideally, you don't want to allow pressure from one line into the line of a lower pressure line. The final pressure should be around | + | Go into lab building, check pressures on Vacmon display. If pressure is higher than 100 torr in any of the lines, pump those lines only one at a time. Open each valve slowly, you will hear the air rush out as the pump pulls the air. Slowly continue to fully open. Open each valve as the pressure is pumped down to the pressure in the next lowest line. Ideally, you don't want to allow pressure from one line into the line of a lower pressure line. The final pressure should be around 1.0 Torr in each line when you leave the lab area if all six lines are opened. The pump will normally bring all 6 lines down to .2-.3 Torr after about an hour. The S1 and S2 lines leak the most during the day and will read the highest in the afternoon when you return, usually between 20-30 Torr in the summer and 15-25 Torr in the winter. Note any unusual vacuum readings as they can indicate a leak beyond what is normal. Sometimes a line isn't pumped down the night before and can read higher than 30 Torr. Pump down the highest ones first as stated above. |
2.2 Filling NIRO with LN2\\ | 2.2 Filling NIRO with LN2\\ | ||
Line 309: | Line 311: | ||
5.2 Opening telescope optics \\ \\ | 5.2 Opening telescope optics \\ \\ | ||
- | When the sun has set and before | + | About 20 minutes after the sun has set, it is dark enough to slew to your first target |
5.3 Check the OPLE carts \\ \\ | 5.3 Check the OPLE carts \\ \\ | ||
Line 331: | Line 333: | ||
Locking Tiptilt using the telescope WFS | Locking Tiptilt using the telescope WFS | ||
+ | |||
+ | [[: | ||
Before locking the star, align the twfs to the red beacon, align the blue beacon to the labao, focus the labao and then focus the twfs as shown before. Turn off the red LED and you are ready to lock the star. | Before locking the star, align the twfs to the red beacon, align the blue beacon to the labao, focus the labao and then focus the twfs as shown before. Turn off the red LED and you are ready to lock the star. | ||
Line 343: | Line 347: | ||
5.6 Locking Lab Tiptilt (only used in special cases) | 5.6 Locking Lab Tiptilt (only used in special cases) | ||
+ | |||
+ | See New Alignment Procedures. | ||
With the star aligned with the acquisition ticks, check to make sure there are counts on the TipTilt server or the white plots in the TT windows have condensed. On the Main Tab of the Telescope GUI, click [TIP/TILT] under the Pointing Servo menu. If this button is not pressed, then the green dots on the tiptilt windows will drift and eventually drop TT lock. The [STAR ACQUIRED] button on Cosmic Debris also starts the Tiptilt servo for all active scopes with one push of a button. On the Tiptilt Servo Control GUI, turn tiptilt [ON] for each telescope or use the [TT ON] button on Cosmic Debris to perform this function for all active scopes. | With the star aligned with the acquisition ticks, check to make sure there are counts on the TipTilt server or the white plots in the TT windows have condensed. On the Main Tab of the Telescope GUI, click [TIP/TILT] under the Pointing Servo menu. If this button is not pressed, then the green dots on the tiptilt windows will drift and eventually drop TT lock. The [STAR ACQUIRED] button on Cosmic Debris also starts the Tiptilt servo for all active scopes with one push of a button. On the Tiptilt Servo Control GUI, turn tiptilt [ON] for each telescope or use the [TT ON] button on Cosmic Debris to perform this function for all active scopes. | ||
Line 392: | Line 398: | ||
Close the OPLE server. | Close the OPLE server. | ||
- | Press the shutdown button on the OPLE System Control gui, wait until all 6 dots are grey, press the Metro button, and wait until it is grey, turn off the metrology laser. | + | Press the shutdown button on the OPLE System Control gui, wait until all 6 dots are grey, press the Metro button, and wait until it is grey. Close the gui on Wazoo. Turn off the metrology laser. |
6.7 Shutting down the Lab \\ \\ | 6.7 Shutting down the Lab \\ \\ | ||
Go to the lab and close the vacuum valves for all telescope lines (valve handle turned perpendicular to the tube). Go outside to the pump shack and close the vacuum valve. Then shut off the vacuum pump. Note any unusual noises or excessive oil temps (above 54 C) reported by the thermometer. Let Craig, Victor or Steve know of any concerns. Lock the padlock. \\ | Go to the lab and close the vacuum valves for all telescope lines (valve handle turned perpendicular to the tube). Go outside to the pump shack and close the vacuum valve. Then shut off the vacuum pump. Note any unusual noises or excessive oil temps (above 54 C) reported by the thermometer. Let Craig, Victor or Steve know of any concerns. Lock the padlock. \\ | ||
- | Go inside the lab with booties: Turn off the silver box, then black box for NIRO and replace the NIRO camera cover if used. Turn off the key for the green alignment laser and turn the power switch off. Turn off Pico 3 controller above the tiptilt camera. Turn off the blue amplifiers for the metrology laser (button is labeled " | + | Go inside the lab with booties: Turn off the silver box, then black box for NIRO and replace the NIRO camera cover if used. Turn off the key for the green alignment laser and turn the power switch off. Turn off Pico 3 controller above the tiptilt camera. Turn off the blue amplifiers for the metrology laser (button is labeled " |
+ | |||
+ | Check the 6 ople computers and RPC's for any components that were not turned off in the OPLESystem Control gui shutdown sequence. Any + signs on the power controllers indicate power was left on. E1 computer is usually on. Push the Off button to turn off the component that is on. Scroll left and right using the UP/DOWN buttons to move to each + sign. Make sure all doors are closed as you leave. | ||