This shows you the differences between two versions of the page.
Both sides previous revision Previous revision Next revision | Previous revision Next revision Both sides next revision | ||
chara:trouble_shooting [2021/11/05 00:44] charaobs [HUT servers] |
chara:trouble_shooting [2021/11/08 17:59] charaobs [Dome does not close] |
||
---|---|---|---|
Line 304: | Line 304: | ||
If a dome will not close after observing, it may need to be closed manually at the bunker. Turn the power to the dome off at the switch, disconnect the cable from the computer control box and connect it to the manual control switches above. Turn on the power to the dome after connecting the cable and use the two controls to close the dropout first, then the slit. The dropout must be fully closed and the slit closed over it to be shut properly. Inform the staff that the dome did not function properly so someone can look at it in the morning. | If a dome will not close after observing, it may need to be closed manually at the bunker. Turn the power to the dome off at the switch, disconnect the cable from the computer control box and connect it to the manual control switches above. Turn on the power to the dome after connecting the cable and use the two controls to close the dropout first, then the slit. The dropout must be fully closed and the slit closed over it to be shut properly. Inform the staff that the dome did not function properly so someone can look at it in the morning. | ||
- | ==== Dome does not close ==== | ||
- | |||
- | \\ | ||
- | If a dome will not close after observing, it may need to be closed manually at the bunker. Turn the power to the dome off at the switch, disconnect the cable from the computer control box and connect it to the manual control switches above. Turn on the power to the dome after connecting the cable and use the two controls to close the dropout first, then the slit. The dropout must be fully closed and the slit closed over it to be shut properly. Inform the staff that the dome did not function properly so someone can look at it in the morning. | ||
====== HUT servers ====== | ====== HUT servers ====== | ||
Line 314: | Line 310: | ||
The HUT servers control functions such as beacon and dichroic movements, heater and dehumidifier usage, and various AO functions. An observer may find that the obsgtk is no longer controlling the beacon LED's, beacon flat or dichroic alignments. This happens on occasion with E2 and other scopes. | The HUT servers control functions such as beacon and dichroic movements, heater and dehumidifier usage, and various AO functions. An observer may find that the obsgtk is no longer controlling the beacon LED's, beacon flat or dichroic alignments. This happens on occasion with E2 and other scopes. | ||
- | ===== Tiptilt Server ===== | + | ===== Tiptilt Server. (These sections apply to the old lab tiptilt system retired in April 2021) ===== |
< | < | ||
Line 320: | Line 316: | ||
</ | </ | ||
+ | |||
==== Tiptilt server complains about the CCD ==== | ==== Tiptilt server complains about the CCD ==== |