Line 217: Line 217:
  
 [[:chara:old_lab_tiptilt_server|Instructions for the old lab tiptilit system are archived here.]] [[:chara:old_lab_tiptilt_server|Instructions for the old lab tiptilit system are archived here.]]
 +
 +===== LabAO =====
 +
 +==== LabAO Shutters ====
 +
 +If the labao shutters are not working, the power button on each of the two red control boxes on the top of the table might need to be pressed. After that the other button needs to be pressed and held for 5 seconds so that the units can be controlled remotely. (Note: look for instructions on table.)
 +
 +==== LabAO DM Electronics ====
 +
 +If a labao server complains that it failed to find the EDAC40 device, try unplugging the power and ethernet cables for DM and then replugging in again.
  