====== New Operating Procedures ====== | ====== New Operating Procedures ====== | ||
- | Last updated 04/17/2024 | + | Last updated 04/22/2024 |
[[: | [[: | ||
Line 168: | Line 168: | ||
- Go to lab building | - Go to lab building | ||
- check pressures on Vacmon display | - check pressures on Vacmon display | ||
- | - To open gui: Chara blue menu, **Vacuum Monitor** | + | - To open gui: Chara blue menu, **Vacuum Monitor, [Status]** |
- S1 and S2 lines leak the most, so open first until they reach the level of the next highest pressure line (usually E1, then W2). Can open all the valves once they are low enough ( ~3 torr or so). | - S1 and S2 lines leak the most, so open first until they reach the level of the next highest pressure line (usually E1, then W2). Can open all the valves once they are low enough ( ~3 torr or so). | ||
- If pressure is higher than 100 torr in any of the lines, pump those lines only one at a time. | - If pressure is higher than 100 torr in any of the lines, pump those lines only one at a time. | ||
+ | |||
==== Turn on Electronics Inside the Lab ==== | ==== Turn on Electronics Inside the Lab ==== | ||
Line 817: | Line 818: | ||
- can use rotation wheel. spica is 90 degrees CW from azimuth of star. | - can use rotation wheel. spica is 90 degrees CW from azimuth of star. | ||
- use steps 222 - 500 and wait for dm to adjust before clicking again. | - use steps 222 - 500 and wait for dm to adjust before clicking again. | ||
- | - as of 10/13, spica requests that telemetry is recorded once data sequence is being recorded. | + | - when finished, on ip_det, press **[FTT SERVO] |
- | - **CD,** | + | |
- Make sure data acquired line comes in on CD, otherwise do **[DATA ACQ]. ** | - Make sure data acquired line comes in on CD, otherwise do **[DATA ACQ]. ** | ||
- or make sure to **[REOPEN]** | - or make sure to **[REOPEN]** | ||
Line 1277: | Line 1277: | ||
- if that doesn' | - if that doesn' | ||
- if it doesn' | - if it doesn' | ||
- | - cart error is jumping about 100 microns every 4-6 seconds. | + | - cart error is jumping about 100 microns every 4-6 seconds. On mircx/ |
- | - On mircx/ | + | |
- | - In the lab, a faint ticking sound could also be heard. | + | |
- | - Turn off the cart | + | - Restart ople server |
- | - On Ople System Controller, restart the ople computer for the cart | + | - Track carts. |
- | - When its done, rehome and check the cart. Issue should go away. | + | |
+ | | ||
+ | - When its done, rehome and check the cart. Issue should go away. | ||
=== Hut Error === | === Hut Error === |