chara:new_operating_procedures [2024/04/24 02:09] bflores [Alignment Sequence for Visible Combiners: SPICA and PAVO] |
chara:new_operating_procedures [2024/04/30 00:04] bflores [Common Issues] |
||
---|---|---|---|
Line 1277: | Line 1277: | ||
- if that doesn' | - if that doesn' | ||
- if it doesn' | - if it doesn' | ||
- | - cart error is jumping about 100 microns every 4-6 seconds. | + | - cart error is jumping about 100 microns every 4-6 seconds. On mircx/ |
- | - On mircx/ | + | |
- | - In the lab, a faint ticking sound could also be heard. | + | |
- | - Turn off the cart | + | - Restart ople server |
- | - On Ople System Controller, restart the ople computer for the cart | + | - Track carts. |
- | - When its done, rehome and check the cart. Issue should go away. | + | |
+ | | ||
+ | - When its done, rehome and check the cart. Issue should go away. | ||
=== Hut Error === | === Hut Error === |