Associated Wikis
(privileges may be required)
Associated Wikis
(privileges may be required)
Next: Introduction, Previous: (dir), Up: (dir) Contents Index
Next: Cooling Down the PAVO Camera, Previous: Top, Up: Top Contents Index
PAVO - Precision Astronomical Visible Observations These notes are designed to be instructions for basic PAVO operation, assuming familiarity with the rest of the CHARA system. The basic observing information, moving telescopes, moving carts, etc is not covered here.
Next: Overview of Alignment Steps, Previous: Introduction, Up: Top Contents Index
The PAVO camera needs to be cooled down at the beginning of the run. Type the following command into the pavo server: > coolon -50
This cools the camera to -50 C. You can read the current temperature from the pavo server. As of 2014, we are doing air cooling (PAVO is no longer hooked up to water cooling). Because of this, the minimum gain setting should be 170 for almost all targets (not 150). If you leave the pavo computer on and the camera plugged-in, then you don’t have to do anything until a PAVO run is over. At the end of the run, type “coolon 0” into the server and it will raise the temperature to 0 Celsius. Once it gets to about 0, it is “safe” to switch anything and everything off without damaging the CCD.
\\ Previous Notes: When PAVO was hooked up to the water cooling, the camera was cooled down to -70 C. [CCD Cool] had to be turned on through the power gui for this to work. The fan is also on, so nothing should break if CCD cool isn’t on; it will just stop at about -55C and not go any lower. The gain numbers in the current mode only apply to this temperature.
Next: Aligning PAVO in the Lab, Previous: Cooling Down the PAVO Camera, Up: Top Contents Index
There are three different phases to the PAVO alignment. An overview of these steps is given below:
Note: The alignment on Beams 1, 2, and 3 are done using VisBeams 1&2 and VisBeams 2&3. All beam pairs should be aligned to the targets on the east-end table (1&2-1, 1&2-2, 2&3-2, 2&3-3).
Next: Description of PAVO Server and GUI, Previous: Overview of Alignment Steps, Up: Top Contents Index
These instructions are for aligning the PAVO optics at the start of a run. After the PAVO beam splitters are put into the path and the PAVO optics are aligned, they shouldn’t need to readjusted for the remainder of the run. The standard CHARA alignment (visbeams, dichroics, M10) should be done on a nightly basis.
Next: Checking Alignment on PAVO Lenses, Previous: Aligning PAVO in the Lab, Up: Aligning PAVO in the Lab Contents Index
Next: Checking Alignment on PAVO Camera, Previous: Basic Alignment of PAVO in the Lab, Up: Aligning PAVO in the Lab Contents Index
Next: Align Wavelength Scale in Lab, Previous: Checking Alignment on PAVO Lenses, Up: Aligning PAVO in the Lab Contents Index
After doing the lab alignment, it is useful to check that the laser light is hitting the PAVO detector. The corner cubes should be in for these steps:
Next: Adjusting PAVO Coarse Alignment, Previous: Checking Alignment on PAVO Camera, Up: Aligning PAVO in the Lab Contents Index
Align wavelength scale of spectrum. The PAVO code is currently set to use the green laser wavelength for the alignment. Experts only for now!
Alignment of the wavelength scale. When the wavelength scale is aligned, the bright band will be about 2 pixels to the left of the center when using the green laser.
Next: Changing focus, Previous: Align Wavelength Scale in Lab, Up: Aligning PAVO in the Lab Contents Index
These are steps to correct severe misalignments in the PAVO system. Do not adjust without prior approval from the PAVO instrument PI.
Next: Step 2 - Correct for offset on L2, Previous: Adjusting PAVO Coarse Alignment, Up: Adjusting PAVO Coarse Alignment Contents Index
Previous: Step 1 - Correct for offset on L1, Up: Adjusting PAVO Coarse Alignment Contents Index
Next: General Alignment Notes, Previous: Adjusting PAVO Coarse Alignment, Up: Aligning PAVO in the Lab Contents Index
If the L1 or M2 mounts have been physically moved on the table during the course alignment, the focus of the beam might need to be adjusted. Do not adjust without prior approval from the PAVO instrument PI.
Previous: Changing focus, Up: Aligning PAVO in the Lab Contents Index
Next: Starting PAVO Software, Previous: Aligning PAVO in the Lab, Up: Top Contents Index
Next: PAVO GUI, Previous: Description of PAVO Server and GUI, Up: Description of PAVO Server and GUI Contents Index
This Section describes the parameters listed on the top of the PAVO server display (see Figure 5.1) and on the PAVO status window.
\\ The top of the server lists a variety of settings and parameters:
Below this panel, provides information on fringe tracking and cart positions:
Next: PAVO Engineering GUI, Previous: PAVO Server/Status Window, Up: Description of PAVO Server and GUI Contents Index
Figure 5.2 shows a screenshot of the PAVO GUI.
The PAVO GUI used to control the instrument. Green indicates that a button is active. In this screenshot, autosave is turned on and the B1, B2, B3 shutters are open. The camera is off; the [GO/STOP] button will turn green when the camera is on. The mask is in (the [MASK OUT] button will turn green when the mask is out.
\\ Detailed description:
PAVO camera display when viewing the image plane on-sky. The three spots show the star light from Beam 1, 2, and 3, respectively from top to bottom. The colored boxes mark the position of each mask hole and are placed by clicking the TARGETS button on the PAVO GUI. The panel on the left shows the image plane with the mask in. The panel on the right shows the image plane with the mask out.
PAVO camera display when viewing the lenslet on-sky. The pupil is segmented by the lenslet array (appearing as vertical bands). Each lenslet band is dispersed by the prism with wavelength going from left to right in each band. The dark spot in the center is the central obscuration from the telescope.
Lab fringes on PAVO in a narrow bandpass. The two panels show the fringes at different frequencies on Beams 1&2 (left) and Beams 2&3 (right).
PAVO waterfall display with fringes on Baseline 2. —- Previous: PAVO GUI, Up: Description of PAVO Server and GUI Contents Index
Figure 5.7 shows the PAVO Engineering GUI which can be opened by clicking the [ENG] button on the PAVO GUI. The Engineering gui is no longer needed as much as it used to be. You may need it to set the mask to the right position at the start of the lab alignment (Basic Alignment of PAVO in the Lab), to adjust the wavelength scale (Align Wavelength Scale in Lab), or to correct for one of the lenses reaching the end of its limit (Large Jump in Image Position). Only change these positions if you are an expert who knows what you are doing. To adjust the position of one of the optics, turn the power for that motor on using the column on the left-hand side. Set the step size using [-] and [+] buttons. Adjust the position using the [<] and [>] buttons. There are pre-defined positions stored in the 0,1,2,3 buttons on the right-hand side. For instance, the mask position is [3] if the mask is in (mask out should be another pre-defined position). For the B1,B2,B3 M1 shutters, 0=open, 1=closed.
\\ ----
Next: On-sky Alignment, Previous: Description of PAVO Server and GUI, Up: Top Contents Index
This section describes the PAVO-specific servers and guis that need to be open at the beginning of the night to observe with PAVO. This list does not include the general CHARA servers that need to be opened by the array operator.
Next: Observing with PAVO, Previous: Starting PAVO Software, Up: Top Contents Index
On-sky alignment of the image and pupil planes. This should be done when switching to a new target/calibrator set in a different part of the sky. You need to re-align after a large telescope slew or more than about 1 hour on target.
• Setting Detector Gain : | ||
• Take Backgrounds: | ||
• Align Image: | ||
• Align Pupil: | ||
• Save Alignment Positions: |
Next: Take Backgrounds, Previous: On-sky Alignment, Up: On-sky Alignment Contents Index
Set the detector gain using the PAVO gui. Confirm that the gain value is correct on the PAVO server.
Next: Align Image, Previous: Setting Detector Gain, Up: On-sky Alignment Contents Index
Next: Align Pupil, Previous: Take Backgrounds, Up: On-sky Alignment Contents Index
Next: Save Alignment Positions, Previous: Align Image, Up: On-sky Alignment Contents Index
Previous: Align Pupil, Up: On-sky Alignment Contents Index
Next: Shutting down PAVO, Previous: On-sky Alignment, Up: Top Contents Index
Next: Target Acquistion, Previous: Observing with PAVO, Up: Observing with PAVO Contents Index
Next: Searching for Fringes, Previous: Important Notes, Up: Observing with PAVO Contents Index
Next: Detailed Observing Notes, Previous: Target Acquistion, Up: Observing with PAVO Contents Index
Previous: Searching for Fringes, Up: Observing with PAVO Contents Index
Next: Recommendations for Setting Beam Order, Previous: Observing with PAVO, Up: Top Contents Index
Compressing the data at the end of the night
Next: Fringe Finding Sources, Previous: Shutting down PAVO, Up: Top Contents Index
These were Mike’s recommendations for setting the W2-S2-W1 beam order (Nov 2008):
Next: Troubleshooting, Previous: Recommendations for Setting Beam Order, Up: Top Contents Index
Next: Random and Important Notes, Previous: Fringe Finding Sources, Up: Top Contents Index
Next: Can't See Laser in Acquistion, Previous: Troubleshooting, Up: Troubleshooting Contents Index
This list outlines the simple things that can prevent you from finding fringes:
If OPLE is not tracking - go to OPLE Server:
Check LDC glass position (possible reason for not finding fringes):
Next: Can't See Starlight on Dectector, Previous: Can't Find Fringes, Up: Troubleshooting Contents Index
Can’t see laser in [ACQ] when aligning TV Tracking on telescope:
Next: Can't See Laser in PAVO Camera, Previous: Can't See Laser in Acquistion, Up: Troubleshooting Contents Index
Check the PAVO server to make sure the gain is set properly. If it is 0, then re-enter the gain setting using the PAVO gui.
The red beacon from the AO board can be seen on the PAVO camera. Check to see if it is visible. If it is not, the starlight is likely not visible as well.
Check if the labao shows starlight when the blue beacon is turned low or off on a bright star, Vmag 3 or brighter. If it is not visible, the dichroic may be misaligned and not sending starlight to the lab. Confirm the dichroic position with the green laser, especially if it is the first night of a run.
If light is still not reaching the detector, then go into the lab, put in the corner cubes, and make sure the laser light is reaching the PAVO table. Look for laser light on the splitters, M1 mirrors above the CLIMB table and L1 lenses facing the PAVO table. Use a white card to see the laser light as it goes into the PAVO table cover slot.
Next: Dispersion is High, Previous: Can't See Starlight on Dectector, Up: Troubleshooting Contents Index
You need to switch the filter to LP02-568 in order to see the red laser in the image plane or the lenslet plane. For the green laser, use filter FB 500-40. Are the corner cubes in? Did the PAVO shutters get saved in the closed position?
Next: Large Jump in Image Position, Previous: Can't See Laser in PAVO Camera, Up: Troubleshooting Contents Index
If the dispersion is high or if you can’t find fringes, you may want to check the glass offsets being sent to the LDC controls by the ople server. High dispersion will produce smeared, asymmetric fringes. This is mostly an issue in terms of tracking on the broadband fringes (post-processing is done using small sub-bandpasses). There are three ways to adjust the glass offsets:
Next: LDCs won't Home, Previous: Dispersion is High, Up: Troubleshooting Contents Index
If there is suddenly a large jump in the position of the image, it might be caused by L2 being at the end of its range. To check:
Next: Using PAVO with Secondary OPLE and Secondary Cosmic Debris, Previous: Large Jump in Image Position, Up: Troubleshooting Contents Index
Next: Homing the lenslet stage (motor 1), Previous: LDCs won't Home, Up: Troubleshooting Contents Index
Next: If you need to home only one of the PAVO motors, Previous: Using PAVO with Secondary OPLE and Secondary Cosmic Debris, Up: Troubleshooting Contents Index
The zaber motor for the lenslet stage can get stuck while homing. This used to hang up the PAVO server while homing motor 1 during the zzero sequence. As of December 2015, Mike updated the PAVO code so that the zzero process skips the homing of motor 1. However, it still moves the stage to 0 and sends it to the last correct position. Therefore, it is important that motor 1 is “homed” and has a realistic 0 position before running the zzero procedure. Here are instructions for how to home the lenslet stage (motor 1):
Next: No light in the pupil plane, Previous: Homing the lenslet stage (motor 1), Up: Troubleshooting Contents Index
Next: PAVO Server - Error communicating with IFW, Previous: If you need to home only one of the PAVO motors, Up: Troubleshooting Contents Index
The pupil location saved in the pavo_zaber_positions file is not currently set to view the pupil. We found that the position of the zaber mount needed to view the pupil will either send the lenslet stage to a position where the imaging lens hits the side of the camera box cover or that the zaber stage extends so far out that it goes to a bad position. When this happens, the zaber stage will lose home and mess up the LENSLET and IMAGE positions. To avoid this we set the pupil position to a location where the mount for the imaging lens blocks all light from hitting the detector (so as not to confuse observers who might accidentally try to align using the pupil position). If the pupil (without the lenslet in) needs to be viewed, then this position will need to adjusted. The last recorded pupil position was 577,549; this would allow one to view about half of the pupil. The blocked pupil position is currently defined as 350,000.
Previous: No light in the pupil plane, Up: Troubleshooting Contents Index
The following error message sometimes appears when starting the PAVO server: “Error communicating with IFW. Could not read from IFW. Is the filter wheel plugged in and on?” Most of the time, restarting the PAVO server will clear this error message and allow PAVO to communicate with the filter wheel. If restarting the server doesn’t help, then check that the cables going into the small, black IFW box on the PAVO table and into the back of the PAVO computer are plugged in securely.
Next: Index, Previous: Troubleshooting, Up: Top Contents Index
Previous: Random and Important Notes, Up: Top Contents Index
Index Entry | Section |
---|